Super Fast Analysis
30 seconds
for a complete gas composition analysis.
Using backflush to detector technology:
- Parallel foreflush and backflush analysis
- Protects the GC column
Dimensions
19″ rack version with mounting brackets and rubber fet
The DynamiQ-R has internal space for custom equipment such as pumps, electronics, filters and sensors.
Contains up to 4 GC ovens
The number and type of column ovens is targeted to the customer application.
Top Level Accurary
500 ppb DL
0.1% RSD
0.01% CV
Designed for stability and robustness. Exact numbers depend on component and application.
Low consumptions
The DynamiQ-R can be configured for minimal carrier gas usage.
- Typical 5 ml/minute carrier gas usage per channel during analysis
- Typical 10 W power consumption per channel
- Whole instrument weight < 20 kg
Easy maintenance
Each instrument is equipped with an easy replaceable cartridge, containing 1 to 4 GC channels with the analytical parts. There are no tubes or capillaries in the instruments which enhances the robustness.
Options
Sample Stream selector
- 3 inlet ports standard
- optional integrated
Depending on the application, the DynamiQ can be equipped with a stream selector which allows to connect multiple sample gases and/or calibration gases. This integrated stream selector switches between the different inlets, so only 1 stream is let into the analyzer at a time. Note that in the case of applying a permanent calibration bottle to one of the inlets, for a standard 3-stream selector, only 2 sample gas stream inlets are possible.
Sample Pump
- optional integrated
- vacuum pump
In case of ambient pressure sample gas(es), 1 or 2 sample vacuum pumps can be integrated.
Multiple carrier Gases
- 1 standard
- up to 4 possible
There is standard one carrier gas port, but the GC analyzer can be configured such that each channel has a different carrier gas.
Data / reporting
- modbus over RS485 or TC/IP
- RS232